SpO2 Accuracy: No Motion: Adults, Pediatrics: 70~100% +/-2% 0~69% unspecified Neonates: 70~100% +/-3% 0~69% unspecified Motion: Adults, Pediatrics: 70~100% +/-3% 0~69% unspecified Neonates: 70~100% +/-4% 0~69% unspecified Low Perfusion: Adults, Pediatrics: 70~100% +/-3% 0~69% unspecified Neonates: 70~100% +/-4% 0~69% unspecified
Pulse Rate Accuracy No Motion: Adults, Pediatrics, Neonates: +/-3% Motion: Adults, Pediatrics, Neonates: +/-5% Low Perfusion: Adults, Pediatrics, Neonates: +/-5%
Displays and Indicators SpO2, Pulse Rate and Pulse Strength Bar Pulse Search Indicator Low Perfusion and Motion Warning Indicator Alarm Silence Indicator AC Power and Battery Charge Indicator Low Battery Warning Indicator
Alarm Audible/Visual Alarms for High/Low SpO2 and PR Audible/Visual Alarms for Low-Battery Alarm Limits Adjustable Alarm Volume Adjustable Pulse Search Indicator Alarm Silence Interval:30,60,90and120sec and OFF Low Perfusion and Motion Warning indicator Sensor Disconnect Alarm Sensor off Alarm |
Non-Invasive Blood Pressure: Measurement type: Adult, Pediatric, Neonatal Measurement range: Systolic 40bpm~270mmHg Diastolic 20bpm~200mmHg Mean 30bpm~220mmHg Accuracy: +/-3 mmHg or 2%(choice max) Resolution: 1mmHg Protection: over pressure
Performance Measurement Range SpO2:0~100% Pulse Rate: 20bpm~250bpm
Environment Operation Temperature: 5~40ºC 41~104ºF Humidity: <=80% Altitude: -390~5,000m(-1,280~16,404ft) Transport/Storage Temperature: -20~+70ºC(-4ºF~158ºF) Humidity: <=95%
Output RS-232 Serial Port for Digital Output Nurse Call Function
Electrical Instrument Power Requirements: 100~240VAC, 50/60Hz, 20VA Fuse Rating: 1.0A
Battery Internal Battery: 12V 3.6Ah Battery Capacity: >=24 hours
Physical Characteristics Dimensions: 264mm(W) x 95mm(H) x 172mm(D) Net Weight: 2.25kg/5lb
Standard EN60601-1/ GB9706.1 EN60601-1-2, EN60601-2-49 |